| Items |
Specification |
| Medical equipment identification number |
219ACBZX00031000 |
| Type |
Digital panoramic/cross-sectional x-ray imaging diagnosis apparatus for use in dental treatment |
| Product name |
PANOURA19 |
| Model |
XP64 |
| High pressure generator |
DC (prior ignition type) |
| X-ray Tube focus |
0.5mm×0.5mm |
| X-ray Tube pressure |
Max. 82kV(58–82kV) |
| X-ray Tube current |
Max. 10mA(2–10mA) |
| Exposure Time |
14 secs,16 secs(Panorama image)
4 secs x 4 (TMJ imaging) |
| Image orientation |
Right-handed rotation |
| Magnification |
1.2 –1.29 (Panorama image, TMJ image) |
| Inherent filtration |
2.5 mm aluminum equivalent or higher |
| Duty cycle |
1:40 |
| Emergency stop method |
Deadman type emergency stop switch |
| Power standards |
Phase 1 (Single Phase)
50–60Hz
Voltage 100VAC
Input 1.5kVA |
| Type of protection against electric shocks |
Class 2 |
| Degree of protection against electric shocks |
Type B machines |
| Weight |
125kg |
| Suitable environment |
Temperature 10–40°C
Humidity 30–75°C (no condensation) |